U3042AM08
Controlling the Test Set with N5222A/B, N5230C, N5232A/B or N5242A/B
Sweep Setup for Multiport and Standalone Modes
When the Test Set is connected to the network analyzer, it is recommended that the analyzer’s
Sweep Setup be configured to Stepped Sweep before calibrating. This is slower than the
Analog Sweep, but is more accurate due to the extra electrical length of the Test Set and test
port cables. Stepped Sweep is available on all models.
1. On the analyzer select STIMULUS > Sweep > Sweep Setup.
2. Select Stepped Sweep.
3. Set the Dwell Time to 5 μsec > OK.
Figure
2
5
Sweep Setup
36
Keysight U3042AM08 User's and Service Guide