Control Lines
Internal Voltage Supply Configuration
The output voltage of pin 12 can be varied from +2 to +5 V. Perform the following
procedure to set the voltage:
1.
Turn
on
the test set.
2.
Measure the voltage between pin 12 and 15 using a multimeter.
3.
Rotate the voltage adjustment trimmer on the front panel until the multimeter
indicates the appropriate voltage
illustrates an example of the connection between the DUT and the test set
without
an external DC power supply. Connect pin 12 to pin 13, for +5 V, and pin 14 to
pin 15 to provide the ground path. Connect each DUT control line to the external
device under test.
CAUTION
You may only connect pin 12–13, and pin 14–15. Damage may result if any other path
is short-circuited.
Figure 4
8
Internal DC Power Configuration (rear panel view)
9
8
1
15
14
13
12
TO DUT (Only required Lines)
GND
12
13
14
15
1
2
8
2 to 5 Vdc
TEST SET
Line 1
Line 2
Line 8
GND
(Rear Panel View)
60