U3042AM08
General Specifications
Environmental Requirements
Refer to the PNA, PNA-L and PNA-X standard documentation for environmental
specifications.
The U3042AM08 complies with all applicable safety and regulatory requirements for
the intended location of use.
• Pressure Altitude (Operation)
3,000 meters (~10,000 feet)
• The instrument can safely operate in a relative humidity of 80% for temperatures
to 31 degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.
Equipment Heating and Cooling
If necessary, install air conditioning and heating to maintain the ambient temperature
within the appropriate range.
CAUTION
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the cabinet. If
the total power dissipated in the cabinet is greater than 800 watts, forced convection
must be used.
Required Conditions for Accuracy Enhanced Measurement
Accuracy–enhanced (error–corrected) measurements require the ambient
temperature of the analyzer and Test Set to be maintained within ± 1 °C of the
ambient temperature at calibration.
Dimensions and Space Requirements
Standard installation of the test set and analyzer is on a customer provided lab
bench, or table top of adequate size and strength. For weight, dimensions and space
requirements, refer to the network analyzer documentation that is used to configure
the Test Set.
CAUTION
The network analyzer is heavy. It is recommended that two individuals, or a
mechanical lift be used to lift or transport the instrument.
Table
4
Instrument Dimension
Model
Weight
Height
Wid th
Depth
U3042M08
11.4 kg (25 lb)
19.1 cm
(7.5 in)
42.5 cm
(16.7 in)
43.2 cm
(17 in)
Keysight U3042AM08 User's and Service Guide
13