U3042AM08
Address and Data Values
Address and Data Values
Setting the Test Port Paths with Address and Data
Example 1:
Example 2:
Table 1
2
Refer to
and
switch paths of the test set.
The address is the first value in the Test Set I/O control or GPIB data command. The
second value controls the source and receiver paths of the ports. To select a test
set port configuration both switches must be set to complete the source and
receiver paths. To do this you must add the data values together to determine the
data command value for each port, which is the second number in the Test Set I/O or
GPIB command.
If the ports have different addresses, two separate address data commands must be
used. Refer to
Port 5 is the Source and Port 8 is the Receiver.
Source Port 5 = address 0, data 1 and Receiver Port 8 = address 16, data 16.
Two separate commands must to be sent, you may use the same dialog box. Send
address 0 and data 1 in one command line, and address 16 and data 16 in the second
line. Refer to
and
.
For further information refer to
“How to Access Interface Control Settings” on page
and
.
If the ports have the same address, only one command is needed.
Port 5 is the Source and Ports 9 is the Receiver.
Source Port 5 = address 0, data 1 and Receiver Port 9 = address 0, data 32. The data
values are added together, the entry will be 0.33.
PNA-L Port Value Select
Address
Source Path
Receiver Path
Data
0
1
2
0
16
32
0
Ports
1
5
9
1
5
9
16
Ports
2
6
10
2
6
10
32
Ports
3
7
11
3
7
11
64
Ports
4
8
12
4
8
12
for in-depth RF path information.