This section will describe how to operate the Test Set with the analyzer.
NOTE
The internal firmware of the analyzer has not been modified for this Test Set option.
Power levels may differ from those indicated on the network analyzer when the Test
Set is connected.
The Test Set is considered a “slave” instrument. A PNA-X or PNA-L must be used to
control the Test Set. There are three methods to control the Test Set. Multiport mode
is recommended due to calibration and ease of use.
•
“Multiport Mode (Option 551)” on page 3
•
•
The U3042AM08 Option 129 (External Source Path Switch) will not operate in
multiport mode (Option 551).When changing from multiport mode to standalone
mode, send commands 0.0, 16.0, 32.0 and 64.0 to ensure the system is set to a
default state.
• Standard - Mechanical switches for low loss. No Option 129 on the serial tag.
• Option 129 - External source path switch ports 1, 5 and 9.
Typeface Key Conventions
The following key conventions are used throughout this document.
•
[HARDKEYS]
are labeled front panel keys
•
SOFTKEYS
are indicated on the instrument display
•
(Italicized in parenthesis)
are menu paths for the 'B' model analyzer
Definitions for Specifications
Specifications describe the warranted performance of calibrated instruments that
have been stored for a minimum of 2 hours within the operating temperature range of
0 to 55 °C, unless otherwise stated, and after a 45 minute warm-up period. Data
represented in this document are specifications unless otherwise noted.
Characteristics
describe product performance that is useful in the application of the
product, but that is not covered by the product warranty. Characteristics are often
referred to as Typical or Nominal values.
•
Nominal
describes representative performance that is useful in the application of
the product when operated over a 20 to 30 °C temperature range. Nominal
performance is not warranted.
•
Typical
describes characteristic performance, which 80% of instruments are
expected to meet when operated over a 20 to 30 °C temperature range. Typical
performance is not warranted.