U3042AM08
General Specifications
Maximum and Recommended Power Levels
CAUTION
It is recommended that you do not operate components near damage levels
(+30 dBm). The power levels must be 3 dB below maximum level to ensure no
damage.
NOTE
Refer to your analyzer’s specifications to determine the maximum input power levels
for the access and test ports, or to optimize the power levels in the receivers.
NOTE
Damage and maximum levels are not necessarily the optimum level.
Table
5
Recommended Power Levels
U3042M08 Test Port RF Power Levels:
PORT 5-12
+27 dBm 0 Vdc
U3042M08 Access Ports:
SOURCE OUT
+20 dBm 0 Vdc
CPLR ARM
+20 dBm 0 Vdc
CPLR THRU
+20 dBm 0 Vdc
RCVR OUT
+20 dBm 0 Vdc
Noise Source IN/OUT
+20 dBm 0 Vdc