U3042AM08
System Setup with N5222A/B and N5242A/B
7. Place the analyzer on top of the Test Set and ensure that the front frame of the
analyzer is positioned slightly forward of the locks that are attached to the Test
Set. Slide the analyzer back so the locks engage the front frame of the analyzer.
8. Secure the analyzer’s lower locking feet to the Test Set’s upper locking feet using
the spring–loaded screws on the locking feet. If the analyzer's lower locking feet
are not aligned with the screw holes in the Test Set’s upper locking feet, loosen
the screws securing the feet to the instrument slightly to align and tighten.
Figure 16
Locking the Analyzer’s
Figure 17
Locking Feet Screws
NOTE
The lock-feet kit (U3021-60002) includes the analyzer and Test Set lock-feet. Refer
to
“Contacting Keysight” on page 9
• PNA - 5023-0132 (Kit includes locking feel and screws)
• Test Set - N5242-20138 (right foot) and N5242-20139 (left foot).
• Screw - 0515-2317
Test Set
Front Frame
Network Analyzer
Front Frame
Slide the network Analyzer
back to engage the lock link
in the back of the network
analyzer's front frame
Lock Link
screw
Slide the network analyzer
back to engage the lock link in
the back of the network
analyzer’s front frame,
Keysight U3042AM08 User's and Service Guide
27