Controlling the Test Set with N5222A/B, N5230C, N5232A/B or N5242A/B
RF Path Configuration with Analyzer Option 029
If the analyzer has Option 029 or H29 (noise figure capability) verify that the path
configuration is set to Default. If it is not in Default configuration, Port 5 will not
operate correctly.
1.
Select
Trace/Chan > Channel
>
Hardware Setup
>
Path Cong...
-
OR-
(Instrument > Setup > Internal Hardware > RF Path Config...)
and in the
drop-
down menu select
Default
>
OK
. Save this configuration as a User Preset by
selecting
Save
>
User Preset
>
Save current state as User Preset
. Do not use the
factory Preset (User Preset Off), the analyzer will return to Option 029 path
configuration.
2.
Verify that the Port 1 Noise Tuner Switch is set to external.
Select
Utilities
System
>
configure
>
Preferences
-
OR-
(Utility > System > System Setup > Preferences > User Preset)
.
If not, select
Meas: Port 1 Noise Tuner Switch
is set to external
.
Figure 3
6
RF Path Configuration
Keysight U3042AM08 User's and Service Guide
43