U3042AM08
Cal Kit Operational Check
Noise Source Out Test
1.
Set the PNA to Standalone Mode.
2.
Remove the jumper from Port 1 CPLR THRU to test set CPLR THRU.
3.
Connect two RF test cables to the PNA Ports 3 and 4. Connect the two cables
together using an adapter.
4.
Perform a response calibration to normalize and remove the loss created the test
cables.
5.
Remove the adapter.
6.
Connect the cable from Port 3 to Noise Source Out.
7.
Connect the cable from Port 4 to the CPLR THRU on the test set.
8.
Send the Test Set I/O command 0.8. Verify that the Ext Noise Source front panel
LED
is On.
9.
Set trace S34 and verify that the trace is approximately -0 to -3 dB.
Figure 5
7
Noise Source Out Trace
72