U3042AM08
Service Information
Troubleshooting the Test Set
If the test set is not operating properly, use the following procedures to isolate and
repair the type of failure encountered. It is recommended that a qualified service
technician perform the following procedures.
Refer to the Keysight PNA Series: Service & Support Home Page at:
To request service, please contact your local service center. In the US, call
800-829-4444. For a listing of service centers worldwide, please visit us at
http://www.keysight.com/find/service
WARNING
No operator serviceable parts inside. Refer servicing to qualified personnel. To
prevent electrical shock do not remove covers.
Non-RF Failures
Typical malfunctions may be the rear panel fan, DUT control lines or front panel
indicator lights.
1.
Verify that the front panel Power Switch is operational.
a.
The rear panel fan and front panel Active LED should operate when the
Standby Switch is in the On position.
2.
AC Line voltage checks (remove AC power from the instrument).
a.
Ensure the proper AC Line voltage is present at the instrument line cord.
b.
Remove the AC power cord from the instrument. Confirm the instrument AC
line module fuse is operational. See
c.
Internal AC line fuses - Remove the bottom cover of the instrument. Near the
rear panel are two fuse holders, remove the fuses and verify that they are
operational. Refer to
.
WARNING
The opening of covers or removal of parts is likely to expose the user to dangerous
voltages. Disconnect the instrument from all voltage sources before being opened.
3. Internal DC Power - During this check you will apply AC power to the instrument
with the top cover removed. Connect the AC power cord to the test set.
a. Set the front panel switch to the Standby position. The fans or indicator lights
should be Off.
b. Set the standby switch to the On position. The rear panel and internal power
supply fans should be operational. The Active LED will be On when test set is
connected to an active the analyzer and is addressed by the analyzer.
c. Verify that the DC LEDs indicators are On.
Keysight U3042AM08 User's and Service Guide
83