3- 14 Keysight N5290/1A Installation Guide
System Installation
PNA, Controller, and Test Head Module Interconnections
3-
N5290/1A Option 202 (Interconnect Kit N5292-60013)
Refer to
.
Figure 3-9
N5290/1A-202—(Interconnect Kit N5292-60013): Install semirigid cables
(N5292-20014 and N5292-20017) and if not done, connect 50 ohm loads
(1810-0118) to the front panel of the PNA and N5292A
IMPORTANT!
The N5227B and N5247B interconnect kits contain 2.4 mm to 3.5 mm
cables. It is important to connect the 2.4 mm end to the PNA and to
connect the 3.5 mm end to the N5292A test set controller. Improper
connections will result in damage to your instruments.
It is important to use two wrenches when connecting the semirigid cables
to the 1/2/3/4 port 1.85 mm outputs front ports of the PNA to avoid
stressing the connectors and causing damage to your PNA (i.e., the
2.4 mm cable end is compatible with the 1.85 mm connector of the PNA).