Installation Note N5222-90114 75
Installation Procedure for the Upgrade
Step 51. Enable Option 405
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must
not
be running.
— Refer to the license message you received from Keysight: Verify that the
analyzer’s model and serial numbers match those on the license message
you received from Keysight.
— A keyboard must be connected to the network analyzer.
Option Enable Procedure
1.
Locate the email(s) from Keysight which contain license file attachments.
These emails are a result of
“License Key Redemption” on page 6
2.
Copy the license file(s) from the email(s) to the
root directory
More than one license file may be copied to the USB flash drive.
3.
Insert the USB flash drive into the PNA’s USB drive slot.
4.
Connect the USB flash drive to the PNA. Within 5 seconds, the PNA
should display a small “New licenses installed” message.
Else, load the license key file(s), manually move your license file(s) to
C:\Program Files\Agilent\licensing
. It may take Keysight License
Manager an extra ~5 seconds to enable the licenses.
5.
Disconnect the USB flash drive from the PNA.
6.
On the analyzer, click or press to open the KLM software from your PNA’s
Windows taskbar by pressing
Start
>
More Programs
>
Keysight License
Manager folder
>
Keysight License
Manager and verify the options are
correct.
For this step, you will need a USB flash drive.
A single license file may contain more than one feature.
A single license file may contain more than one feature.
Attempting to re-install a license file that is already installed may generate
a “Corrupt Media” error message. Ignore this message.