64 Installation Note N5222-90114
Installation Procedure for the Upgrade
-
Connect New Port 4 N5240-60098 (x1) and Verify Port 2 N5240-60097
Blue Bias-Tee Combiner Cables and Clamps from the A73 and A74 Bias
Tees “RF-IN” to the A70 LFE Board “Port2” and “Port4” Connectors and
Install Clamps onto the Ferrite Beads
3.
Install the N5240-60097 (x1) and N5240-60098 (x1) cable as shown. Note
the orientation of the cable (item
through
).
Figure 37
Connect new N5240-60098 from Port 4 and verify N5240-60097 from Port 2
Bias Tee Combiners to A70 LFE Board and Install Clamps onto the Ferrite
Beads (N5240-60098 (x1), and 1400-1391 (x3))
Verify Port
2 cable is
properly
routed and
secured.