Keysight B1542A User’s Guide, Edition 10
5-7
PLSDIV Test Definitions
PLSDIV DC IdVd
PLSDIV DC IdVd
This test definition controls SMUs to apply the drain voltage and the gate voltage,
measures the drain current, and displays the Id-Vd characteristics.
Device Parameters
The following parameters are available to set the test condition.
Polarity
Polarity of source output. Nch (applies the specified value) or
Pch (applies the negative specified value).
Lg
Gate length, in m.
1 nm, 4 digits or 1 nm resolution.
Wg
Gate width, in m.
1 nm, 4 digits or 1 nm resolution.
Temp
Temperature, in degree. 4 digits or 0.001 degree resolution.
IdMax
Maximum drain current measurement range, in A. 0 to 80 mA,
4 digits or 1
A resolution.
Test Parameters
The following parameters are available to set the source output.
CableR
Residual resistance of the gate measurement path from SMU to
terminator.
0
, 4 digits or 1 m
resolution.
TermR
Resistance of terminator.
1 m
, 4 digits or 1 m
resolution.
TermR=No applies 1E+30
.
Gate
SMU connected to the gate terminal.
VgStart
Secondary sweep start voltage for the gate terminal.
-
10 to 10
V, 4 digits or 10 mV resolution.
VgStop
Secondary sweep stop voltage for the gate terminal.
-
10 to 10
V, 4 digits or 10 mV resolution.
VgStep
Secondary sweep step voltage for the gate terminal.
-
10 to 10 V,
5 digits or 1 mV resolution.
Drain
SMU connected to the drain terminal.
VdStart
Primary sweep start voltage for the drain terminal.
-
10 to 10 V,
4 digits or 10 mV resolution.
VdStop
Primary sweep stop voltage for the drain terminal.
-
10 to 10 V,
4 digits or 10 mV resolution.
VdStep
Primary sweep step voltage for the drain terminal.
-
10 to 10 V,
5 digits or 1 mV resolution.
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......