Keysight B1542A User’s Guide, Edition 10
2-9
Installation
Hardware Installation
Figure 2-5
Pulsed IV Test System Connection Overview
Scope
DC source monitor
Triaxial cables
Pulse generator
D-sub 9 pin cables
Viking cable
to Drain
to Gate
Switch controller
Switch control
distributor
to SMU Force
for gate voltage
at DC test
to SMU Force
for drain voltage
at DC test
to SMU
Force
for drain
DC bias
Drain pulse/dc switch unit
Gate pulse/dc switch unit
SMA cable, 0.3 m
SMA cable, 0.3 m
SMA cable, 1.5 m
DUT cable, 1.5 m
to input
channel
for drain
to input
channel
for gate
Drain Current Monitor
Gate Pulse Monitor
SMU
SMU
BNC cable
to TRIGGER OUT
to AUX Trig In
to output channel
for gate pulse
Terminator
DUT cable, 1.5 m
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......