3-12
Keysight B1542A User’s Guide, Edition 10
Performing System Setup and Compensation
Executing Action
Executing Action
Execute the action as shown below. Before starting the action, confirm that all
system components are connected to the system controller and available on the same
GPIB bus and on the same LAN for the DSO/MSO 9000 series or S series
oscilloscope.
1. Open the PLSDIV Setup test setup from your preset group.
2. Set the DcSwitch field to Yes. Do not set No.
3. Click the Extended Setup button and set the SWAddress value.
SWAddress
GPIB address of the switch controller. Integer, 1 to 32.
4. Select the action by using the Action field.
•
System Configuration
“System Configuration” on page 3-10
•
Compensation
“Compensation Action” on page 3-13
•
Skew Measurement
•
Drain Cable Replacement
•
Pgu Compensation
“Pgu Compensation Action” on page 3-13
.
5. Click the Single button at the upper right corner of the EasyEXPERT screen or
the Desktop EasyEXPERT screen.
6. Follow the pop-up window and dialog box to complete the action.
You will get navigation for performing the system setup and completing the
action.
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......