Keysight B1542A User’s Guide, Edition 10
3-13
Performing System Setup and Compensation
Executing Action
Compensation Action
Before starting the Compensation action, connect triaxial cables as follows.
1. Connect a triaxial cable to the Force connector of the SMU specified by the
Drain Bias SMU CH
field on the Configuration dialog box or the SMU specified
by the
Force SMU (Slot No.)
field on the SMU CH for Compensation dialog box
if you set.
2. Connect a triaxial cable to the Sense connector of the SMU specified by the
Drain Bias SMU CH
field on the Configuration dialog box or the SMU specified
by the
Sense SMU (Slot No.)
field on the SMU CH for Compensation dialog box
if you set.
If you are using the 4155B/C, connect the cable to the Force connector of the
SMU specified by the
Sense SMU (Slot No.)
field on the SMU CH for
Compensation dialog box.
Pgu Compensation Action
Previously, the Compensation action must be completed successfully.
Before starting the Pgu Compensation action, click the Extended Setup button and
set the following parameters. They must be the values you are going to set when
performing the pulsed IV measurements actually.
PulseBase
PGU output baseline voltage,
-
3.0 to 3.0 V
TransTime
Pulse leading/trailing edge transition time, 2 ns to 500 ns
MeasTime
Measurement timing at the pulse top,
-
1 to 1
For the setup parameters, see
.
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......