Keysight B1542A User’s Guide, Edition 10
5-5
PLSDIV Test Definitions
PLSDIV Capt Wave
Test Parameters
The following parameters are available to set the source output.
PulseWidth
Pulse width of PGU output pulse. 10 ns to 1
s, 4 digits or 1 ns
resolution. See
.
Vg
Voltage applied to the gate terminal. PGU output.
-
4.5 V to
4.5 V, 4 digits or 10 mV resolution.
PulseBase
PGU output baseline voltage.
-
3.0 to 3.0 V, 4 digits or 10 mV
resolution. See
. Pulse amplitude
must be > 55 mV and < 4.5 V. To use the gate voltage automatic
adjustment function, set the pulse amplitude to < 4.0 V.
Vd
Voltage applied to the drain terminal. SMU output.
-
10 V to
10 V, 5 digits or 1 mV resolution.
Extended Test
Parameters
The following parameters are available to set the test condition. For more details,
see
TransTime
Pulse leading/trailing edge transition time, in s. 2 ns to 500 ns, 3
digits or 1 ns resolution.
For the 81101A, minimum value is 5 ns.
For the 81150A, minimum value is 2.5 ns when the following
conditions are satisfied, or 7.5 ns.
•
Pulse amplitude is 2.5 V or less.
•
Both pulse base and top are within
2.5 V.
MeasTime
Measurement timing at the pulse top. 0 to 1, 3 digits or 0.01
resolution.
SmplNum
Oscilloscope sampling number. 16 to 262144, 6 digits or 1 step.
SmthNum
Oscilloscope smoothing number. 3 to 4001, 4 digits or 1 step.
SmthStat
Status of the oscilloscope smoothing. Enable or Disable.
AvgNum
Oscilloscope averaging number. 1 to 4095, 4 digits or 1 step.
VdAdjust
Status of the drain voltage automatic adjustment. Enable or
Disable.
VgAdjust
Status of the gate voltage automatic adjustment. Enable or
Disable.
ConnectChk
Status of the connection check at initialization. Enable or
Disable.
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......