2-18
Keysight B1542A User’s Guide, Edition 10
Installation
Software Installation
After Software Installation
1. Launch EasyEXPERT or Desktop EasyEXPERT. And click the Start
EasyEXPERT button.
2. Open your workspace or create a new workspace.
3. Display the Application Test tab screen.
4. Import the pulsed IV test definitions listed below from the following folder.
<program folder>\Agilent\PLSDIV\TestDefinitions\Import1st
•
PLSDIV IdVd
•
PLSDIV IdVd [2]
•
PLSDIV IdVg
•
PLSDIV IdVg [2]
•
PLSDIV DC IdVd
•
PLSDIV DC IdVg
•
PLSDIV Capt Wave
•
PLSDIV Setup
•
PLSDIV Reset
•
PLSDIV IV SMU
5. Import the pulsed IV test definitions listed below from the following folder.
<program folder>\Agilent\PLSDIV\TestDefinitions\Import2nd
•
PLSDIV DC IdVd SMU
•
PLSDIV DC IdVg SMU
6. Open the PLSDIV Setup test definition and perform the system setup as
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......