Keysight B1542A User’s Guide, Edition 10
5-17
PLSDIV Test Definitions
PLSDIV DC IdVg SMU
VdStop
Secondary sweep stop voltage for the drain terminal.
-
10 to 10
V, 4 digits or 10 mV resolution.
VdStep
Secondary sweep step voltage for the drain terminal.
-
10 to 10
V, 5 digits or 1 mV resolution.
•
VgStart, VgStop, VgStep
The parameters set the primary sweep source used to apply the gate voltage. The
number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If VgStep=0, the measurement voltage is
VgStart only.
Number of sweep points = (VgStop
-
VgStart) / 1
•
VdStart, VdStop, VdStep
The parameters set the secondary sweep source used to apply the drain voltage.
The number of sweep points is given by the following formula. Then ignore the
numbers after the decimal point. If VdStep=0, the measurement voltage is
VdStart only.
Number of sweep points = (VdStop
-
VdStart) / 1
•
Gate voltage
To apply the specified voltage (Vg) to the gate terminal, the SMU output voltage
(VgSet) will be set to the following value.
VgSet = Vg * (TermR+CableR) / TermR
Extended Test
Parameters
The following parameters are available to set the test condition.
IgLimit
Gate current compliance, in A. 1 nA to 100 mA, 3 digits or 1 nA
resolution. Set the value that covers current to the terminator.
For example, the current is 100 mA for TermR=50
with
Vg=5 V.
Ranging
Current measurement ranging mode. Auto, LimitedAuto, or
Fixed.
Irange
Current measurement range. 10 pA to 100 mA. 6 digits or 10
pA resolution. Available measurement range depends on the
type of SMU.
ADCType
Only for the E5260/E5270 series. Type of A/D converter.
HighSpeed or HighResolution
IntegMode
Only for the E5260/E5270 series. Integration mode. Auto,
Manual, or PLC.
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......