Keysight B1542A User’s Guide, Edition 10
6-15
PLSDIV TIS Commands
Command Reference
Auto Adjustment
This function is set by the PlsdivSetAdjustStat command, and is used to adjust the
gate voltage automatically so that the
vg_mon
value is close to the
vg_high
value
specified by the PlsdivSetVgLevel command.
Response
Status code. Numeric value. To read the error message, use PlsdivErrorMessage. If
no error is detected by this command execution, 0 is returned.
PlsdivInit
This command initializes the pulsed IV test system. This command must be
executed before any other Plsdiv command.
Filename
PlsdivInit.exe
Write Type
String
Read Type
String. However, this command does not return any output parameter.
Write String
This command needs the following input parameters in this sequence.
connect_check
Connection check status. Numeric value, 0 (does not perform)
or 1 (performs connection check in the PlsdivInit process).
gate_zero
Gate zero status. Numeric value, 0 (does not perform) or
1 (performs gate monitor channel offset measurement in the
PlsdivInit process).
Gate Zero
This function is used to measure the offset data of the gate voltage monitor path and
update the offset value in a system file. The offset value is used until the next
execution. Before performing the offset measurement, set up the test system
measurement path properly and open the measurement terminals.
Connection Check
This function is used to confirm if the divider, terminator, and bias-T are connected
properly. For the proper connections, see
. The connection
check can detect the following improper connections.
•
divider is not connected
•
terminator is not connected
•
T-adapter is connected instead of divider
•
bias-T’s DC+RF connector is connected to the gate monitor channel
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......