Keysight B1542A User’s Guide, Edition 10
3-7
Performing System Setup and Compensation
Starting System Setup
Figure 3-1
Keysight Connection Expert
If You Use DSO/MSO 9000 Series or S Series
Oscilloscope
Perform the following procedure to know the IP address of the DSO/MSO 9000
series or S series oscilloscope and set it to Keysight Connection Expert. It must be
also set to the system setup configuration (see IP Address in
). Before
performing this procedure, the oscilloscope and B1500A (or computer) must be
connected to the LAN of the same subnet mask.
1. On the oscilloscope front panel, check the IP address of the oscilloscope. See the
oscilloscope manual for checking the IP address (web address).
2. If the B1500A (or computer) displays the Start EasyEXPERT window, close it
first.
3. Launch Keysight Connection Expert.
4. Highlight
LAN(TCPIP0)
.
5. Add the oscilloscope to the LAN instruments. Then use the IP address checked
at the step 1. Also the Device name must be
inst0
.
6. Close Keysight Connection Expert.
GPIB0
System Controller
VISA interface ID
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......