Keysight B1542A User’s Guide, Edition 10
3-9
Performing System Setup and Compensation
Starting System Setup
8. Save the test setup to your preset group (My Favorite Setup).
Table 3-4
Desktop EasyEXPERT Execution Mode
Figure 3-2
Keysight Connection Expert
Mode
Field name
Setting value or description
Offline
a
a. Even though the execution mode is
Offline
, the pulsed IV system soft-
ware can control the DC source monitor.
Select this mode for using the 4155B, 4155C, 4156B, 4156C,
E5260A (or E5262A/E5263A), or E5270B.
Model
Select 4155B, 4155C, 4156B, or 4156C to be
used. Or select B1500A for using the E5260A
(or E5262A/E5263A) or E5270B.
Online
Select this mode only for using the B1500A.
VISA interface ID
VISA interface ID of the computer’s GPIB
interface. It must be
GPIB0
.
GPIB address
Set the GPIB address of the B1500A.
GPIB0
VISA interface ID
System Controller
GPIB address
hostname1
4156C (GPIB0::17::INSTR)
4156C (GPIB0::17::INSTR)
GPIB address of instrument
(ex. 4156C)
Содержание B1542A Pulsed IV Package
Страница 1: ...Keysight Technologies B1542A Pulsed IV Package for B1500A EasyEXPERT User s Guide ...
Страница 9: ...1 Introduction ...
Страница 18: ...1 10 Keysight B1542A User s Guide Edition 10 Introduction Accessories and Options ...
Страница 19: ...2 Installation ...
Страница 41: ...3 Performing System Setup and Compensation ...
Страница 54: ...3 14 Keysight B1542A User s Guide Edition 10 Performing System Setup and Compensation Executing Action ...
Страница 55: ...4 Performing Measurement ...
Страница 68: ...4 14 Keysight B1542A User s Guide Edition 10 Performing Measurement DC I V Measurements ...
Страница 69: ...5 PLSDIV Test Definitions ...
Страница 108: ...5 40 Keysight B1542A User s Guide Edition 10 PLSDIV Test Definitions Parameters ...
Страница 109: ...6 PLSDIV TIS Commands ...
Страница 133: ...7 Status Code and Error Messages ...
Страница 159: ......