5-8
Delta, Pulse Delta, and Differential Conductance
Model 6220/6221 Reference Manual
Return to
Figure 5-4
System connections – PC control of Model 622x
DUT test connections
WARNING
Before making or breaking test connections, the Models 622x
and 2182/2182A must be turned off and the line cords must be
disconnected from AC line power.
DUT test connections are shown in
. This connection scheme uses
guarding and an earth grounded test fixture that is equipped with an interlock
switch.
Guarding
The purpose of guarding is to eliminate the effects of leakage current and cable
capacitance that exist between output high and output low. In the absence of
guard, leakage and cable capacitance can adversely affect the performance of the
Model 622x. Guarding can be used for all DUT testing and is most effective when
sourcing low current and/or testing high impedance DUT.
In order to use guarding as shown in
, the driven guard must be con-
nected to the inner shield of the OUTPUT connector of the Model 622x. Use the
following procedure to connect the inner shield to Guard.
ETHERNET
iEEE-488
RS-232
TRIGGER LINK
RS-232
iEEE-488
TRIGGER
LINK
8501 Trigger Link Cable
PC
OR
IEEE-488
Cable
Ethernet Cross-over Cable
(RJ-45 male/male)
(6221only)
Keithley 622x Current Source
Keithley 2182/2182A Nanovoltmeter
(null-modem, male-to-male)
RS-232 Cable
6220: GPIB selected
6221: GPIB or Ethernet selected
RS-232 On
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...