2-22
Output Connections
Model 6220/6221 Reference Manual
Return to
(B)
Three 5-way binding posts, or one 3-lug female triax connector
–
•
If using the supplied triax cable (terminated with alligator clips), mount three
5-way binding points to the test fixture.
•
If using a triax cable that is terminated with 3-slot male triax connectors on
both ends (e.g., Model 7078-TRX-3), mount a 3-lug female triax bulkhead
connector to the test fixture.
Connect the triax cable to the Model 622x
OUTPUT
.
(C)
Banana jacks
– Output Low and/or banana jack Guard can be accessed at
the
LO
and
GUARD
banana jacks on the Model 622x. Use banana plug
cables to make these connections.
(D)
Chassis ground
– the chassis of the test fixture must be connected to chas-
sis ground of the Model 622x. For the Guard triax configuration, the chassis of
the test fixture can be connected to the chassis ground screw on the rear
panel of the Model 622x. This connection to the test fixture could be a direct
soldered connection, a banana plug connection, or a screw terminal connec-
tion.
(E)
Binding posts
– 5-way copper binding posts will accommodate cables termi-
nated with banana plugs or spade lugs. The cable supplied with the Model
2182/2182A is terminated with copper spade lugs. A clean copper-to-copper
connection minimizes thermal EMFs.
(F)
Safety earth ground
– the test fixture must have a screw terminal that is
used exclusively for connection to safety earth ground. See the “
” WARNING before for more information on this connection.
Test fixture chassis
•
The chassis of the test fixture must be metal so it can function as a safety
shield.
•
The test box must have a lid that closes to prevent contact with live circuitry.
Make sure to install an interlock as previously described.
•
The test fixture must have a screw terminal that is used exclusively for con-
nection to safety earth ground.
WARNING
To provide protection from shock hazards, the test fixture chas-
sis must be properly connected to safety earth ground. A
grounding wire (#18 AWG or larger) must be attached securely
to the test fixture at a screw terminal designed for safety
grounding. The other end of the ground wire must be attached
to a known safety Earth Ground.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...