Model 6220/6221 Reference Manual
Calibration 17-17
Return to
2
μ
A
2
μ
A
0
μ
A
-2
μ
A
SOUR:CURR:RANG 2e-6
SOUR:CURR 2e-6
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-6
CAL:PROT:SOUR <-FS_DMM_Reading>
200nA
200nA
0nA
-200nA
SOUR:CURR:RANG 2e-7
SOUR:CURR 2e-7
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-7
CAL:PROT:SOUR <-FS_DMM_Reading>
OUTP OFF
20nA
20nA
0nA
-20nA
OUTP ON
SOUR:CURR:RANG 2e-8
SOUR:CURR 2e-8
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-8
CAL:PROT:SOUR <-FS_DMM_Reading>
2nA
2nA
0nA
-2nA
SOUR:CURR:RANG 2e-9
SOUR:CURR 2e-9
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-9
CAL:PROT:SOUR <-FS_DMM_Reading>
OUTP OFF
Table 17-4 (cont.)
Remote current calibration summary
Range
Currents
Commands*
* Use DMM (
Figure 17-1
) for 100mA-200nA ranges.
* Use picoammeter (
Figure 17-2
) for 20nA and 2nA ranges.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...