Model 6220/6221 Reference Manual
Delta, Pulse Delta, and Differential Conductance
5-63
Return to
7.
When finished with Differential Conductance, it can be disarmed by send-
ing the following command:
8.
Recall stored Differential Conductance readings
– Model 2182/2182A
Differential Conductance readings were sent to the buffer of the Model
622x. Send the following read command to read the buffer:
NOTE
If a read command is sent when Differential Con-
ductance is not running, error -221 Settings Con-
flict will occur. See
for details on errors
associated with Differential Conductance opera-
tion.
NOTE
Details on using Model 622x “
”
are provided on
.
SOUR:SWE:ABOR
‘
‘
Stops Delta and places the Model 2182/2182A in
the local mode.
TRACe:DATA?
‘
‘
Read
Differential Conductance readings stored in
622x buffer.