11-12 Status Structure
Model 6220/6221 Reference Manual
Figure 11-4
Standard event status
Operation event status
The used bits of the operation event register (
) are described as
follows:
•
Bit B0, Calibrating (CAL)
— Set bit indicates that the Model 622x is
calibrating.
Error/status code: +121
•
Bit B1, Sweep Done (SwpD)
— Set bit indicates that a sweep, Delta,
Differential Conductance, or a pulse sweep for Pulse Delta is finished.
Error/status code: +122
•
Bit B2, Sweep Aborted (SwpA)
— Set bit indicates that a sweep, Delta,
Differential Conductance, or a pulse sweep for Pulse Delta has been
aborted.
Error/status code: +123
* ESR ?
PON
(B7)
URQ
(B6)
CME
(B5)
EXE
(B4)
DDE
(B3)
QYE
(B2)
(B1)
(B0)
OR
Standard Event
Status Register
Standard Event
Status Enable
Register
PON = Power On
URQ = User Request
CME = Command Error
EXE = Execution Error
DSE = Device-Specific Error
OPC
PON
(B7)
URQ
(B6)
CME
(B5)
EXE
(B4)
DDE
(B3)
QYE
(B2)
(B1)
(B0)
OPC
* ESE
* ESE ?
(B15 - B8)
(B15 - B8)
To Event
Summary
Bit (ESB) of
Status Byte
Register.
&
&
&
&
&
&
&
QYE = Query Error
OPC = Operation Complete
& = Logical AND
OR = Logical OR
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...