Model 6220/6221 Reference Manual
Calibration 17-23
Return to
Table 17-8
CAL:PROT:GUAR command parameter ranges
Step
Parameter
STEP1
0.00 to +50.0e-3
STEP2
-50.0e-3 to 0.0
NOTE: Step1 and Step 2 may be performed in
any order, but both must be completed, and
each step must be initialized with INIT before
being run.
Table 17-9
Calibration errors
Number
Description
+500
"Date of calibration not set"
+501
"Next date of calibration not set"
+502
"Calibration data invalid"
+503
"DAC calibration overflow"
+504
"DAC calibration underflow"
+505
"Source offset data invalid"
+506
"Source gain data invalid"
+507
"Sense offset data invalid"
+508
"Sense gain data invalid"
+509
"Not permitted with cal locked"
+510
"Not permitted with cal un-locked"
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com