Model 6220/6221 Reference Manual
Wave Functions (6221 Only)
7-25
Return to
Programming arbitrary waveforms
1.
Restore defaults with this command:
*RST
2.
Configure the waveform.
Examples
– The following commands configure 100kHz arbitrary wave
with an amplitude of 25mA, 0mA offset, phase marker off, and 20 second
duration:
3.
Arm and trigger the waveform, turn on output:
4.
To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:ARB:DATA -0.5, -0.25,
0, 0.3, 0.4
‘
‘
Define arbitrary data points in
location 0.
SOUR:WAVE:ARB:COPY 1
‘ Copy points to location 1.
SOUR:WAVE:FUNC ARB1
‘ Select arb wave, location 1.
SOUR:WAVE:FREQ 1e5
‘ Set frequency to 100kHz.
SOUR:WAVE:AMPL 25e-3
‘ Set amplitude to 25mA.
SOUR:WAVE:OFFS 0
‘ Set offset to 0.
SOUR:WAVE:PMAR:STAT OFF
‘ Turn off phase marker.
SOUR:WAVE:DUR:TIME 20
‘ 20 second duration.
SOUR:WAVE:RANG BEST
‘ Select best fixed source range.
SOUR:WAVE:ARM
‘ Arm waveform.
SOUR:WAVE:INIT
‘ Turn on output, trigger waveform.
SOUR:WAVE:ABOR
‘ Stop generating waveform.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...