17-16 Calibration
Model 6220/6221 Reference Manual
Return to
Table 17-4
Remote current calibration summary
Range
Currents
Commands*
100mA
100mA
0mA
-100mA
OUTP ON
SOUR:CURR:RANG 0.1
SOUR:CURR 0.1
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -0.1
CAL:PROT:SOUR <-FS_DMM_Reading>
20mA
20mA
0mA
-20mA
SOUR:CURR:RANG 2e-2
SOUR:CURR 2e-2
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-2
CAL:PROT:SOUR <-FS_DMM_Reading>
2mA
2mA
0mA
-2mA
SOUR:CURR:RANG 2e-3
SOUR:CURR 2e-3
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-3
CAL:PROT:SOUR <-FS_DMM_Reading>
200
μ
A
200
μ
A
0
μ
A
-200
μ
A
SOUR:CURR:RANG 2e-4
SOUR:CURR 2e-4
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-4
CAL:PROT:SOUR <-FS_DMM_Reading>
20
μ
A
20
μ
A
0
μ
A
-20
μ
A
SOUR:CURR:RANG 2e-5
SOUR:CURR 2e-5
CAL:PROT:SOUR <+FS_DMM_Reading>
SOUR:CURR 0
CAL:PROT:SOUR <0_DMM_Reading>
SOUR:CURR -2e-5
CAL:PROT:SOUR <-FS_DMM_Reading>
* Use DMM (
) for 100mA-200nA ranges.
) for 20nA and 2nA ranges.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...