Model 6220/6221 Reference Manual
Calibration
17-5
Return to
Calibration summary and considerations
Calibration summary
Model 622x calibration procedures include:
•
Front panel calibration,
•
Calibration considerations
When performing the procedures:
•
Be sure to restore factory defaults as outlined above.
•
Make sure that the test equipment is properly warmed up and connected to
the correct Model 622x connector.
•
Be sure that a shorting connector is connected to the INTERLOCK connec-
tor, or you will not be able to turn on the output for calibration. See Section
2 for details on the interlock.
•
Make sure the test equipment and Model 622x calibration settings are
correct.
Calibration procedures
Front panel calibration
Using the calibration menu
summarizes the main calibration menu. To enter the menu, press the
MENU key, select CAL, then press ENTER. Note that the calibration menus for the
Models 6220 and 6221 are almost identical except that the Model 6221 menu has
a WAVE calibration selection to calibrate the waveform generator while the Model
6220 does not.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...