Calibration Reference
B-11
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose
To program the 100k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step8 <Cal_resistance>
Parameter
<Cal_resistance> = 90E3 to 110E3 [
Ω
]
Description
:STEP8
programs the 100k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP8 100E3
Program 100k
Ω
step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose
To program the 1M
Ω
comprehensive calibration step.
Format
:cal:prot:dc:step9 <Cal_resistance>
Parameter
<Cal_resistance> = 900E3 to 1.1E6 [
Ω
]
Description
:STEP9
programs the 1M
Ω
comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use
the 1E6 value whenever possible, or the closest possible value.
Example
:CAL:PROT:DC:STEP9
Program 1M
Ω
calibration step.
:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose
To program the 10mA comprehensive calibration step.
Format
:cal:prot:dc:step10 <Cal_current>
Parameter
<Cal_current> = 9E-3 to 11E-3 [A]
Description
:STEP10
programs the 10mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 9E-3 to 11E-3. Use
the 10E-3 value whenever possible for best results.
Example
:CAL:PROT:DC:STEP10 10E-3
Program 10mA step.
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...