Performance Verification
1-7
Test considerations
When performing the verification procedures:
•
Be sure to restore factory defaults as outlined above.
•
Make sure that the equipment is properly warmed up and connected to the correct input
terminals. Also make sure that the INPUTS switch is in the correct position.
•
Do not use autoranging for any verification tests because autorange hysteresis may
cause the Model 2700 to be on an incorrect range. For each test signal, you must manu-
ally set the correct range for the Model 2700 using the range keys.
•
Make sure the calibrator is in operate before you verify each measurement.
•
Always let the source signal settle before taking a reading.
WARNING
Observe the following safety precautions when performing these tests:
•
Some of the procedures in this section may expose you to dangerous
voltages. Use standard safety precautions when such dangerous volt-
ages are encountered to avoid personal injury or death caused by elec-
tric shock.
•
For the front panel terminals only, the maximum common-mode volt-
age (voltage between INPUT LO and chassis ground) is 500V peak.
Exceeding this value may cause a breakdown in insulation, creating a
shock hazard.
•
For the plug-in modules, the maximum common-mode voltage (volt-
age between any plug-in module terminal and chassis ground) is 300V
DC or 300V RMS. Exceeding this value may cause a breakdown in
insulation, creating a shock hazard.
•
When using the front panel terminals simultaneously with plug-in
modules, all cable insulation voltage ratings must equal or exceed the
maximum voltage applied to either the front panel terminals or the
plug-in module terminals.
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...