Calibration Reference
B-7
:DATE
(:CALibration:PROTected:DATE)
Purpose
To send the calibration date to the instrument.
Format
:cal:prot:date <year>, <month>, <day>
Parameter
<year> = 1999 to 2098
<month> = 1 to 12
<day> = 1 to 31
Query format
:cal:prot:date?
Response
<year>, <month>, <day>
Description
The
:DATE
command allows you to store the calibration date in instrument
memory for future reference. You can read back the date from the instru-
ment over the bus by using the
:DATE?
query or the CALIBRATION
selection in the front panel CAL menu.
NOTE
The year, month, and day parameters must be delimited by commas.
Example
:CAL:PROT:DATE 1999,12,16
Send cal date (12/16/1999).
:CAL:PROT:DATE?
Request cal date.
:NDUE
:CALibration:PROTected:NDUE)
Purpose
To send the next calibration due date to the instrument.
Format
:cal:prot:ndue <year>, <month>, <day>
Parameter
<year> = 1999 to 2098
<month> = 1 to 12
<day> = 1 to 31
Query format
:cal:prot:ndue?
Response
<year>, <month>, <day>
Description
The
:NDUE
command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date from the
instrument over the bus by using the
:NDUE?
query or the front panel CAL
menu.
NOTE
The next due date parameters must be delimited by commas.
Example
:CAL:PROT:NDUE 2000,12,16
Send due date (12/16/2000).
:CAL:PROT:NDUE?
Request due date.
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...