![Keithley 2700 Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/keithley/2700/2700_service-manual_3944616071.webp)
Calibration
2-23
5.
Press SHIFT then TEST, then display TEST:CALIB with the up or down range key.
Press ENTER, select RUN, then enter the appropriate calibration code (default:
002700).
6.
Using the up or down range key, select CARD at the CAL:RUN prompt, then press
ENTER.
7.
Set the display value to the cold calibration temperature (˚C) you measured in Step 3,
then press ENTER to complete Model 7700 calibration.
Remote Model 7700 calibration
1.
Connect the Model 7700 to the 7797 Calibration/Extender Board (see “Extender board
connections” above).
2.
With the power off, install the Model 7700/7797 combination in Slot 1, and select the
rear inputs with the INPUTS switch. Allow three minutes for thermal equilibrium.
3.
Accurately measure and record the cold temperature of the Model 7700 card surface at
the center of the card.
4.
Press in and hold the Model 2700 OPEN key while turning on the power.
5.
Enable calibration by sending the :CODE command. For example, the default com-
mand is:
:CAL:PROT:CODE 'KI002700'
6.
Initiate calibration by sending the following command:
:CAL:PROT:CARD1:INIT
7.
Calibrate the Model 7700 with the following command:
:CAL:PROT:CARD1:STEP0 <temp>
Here <temp> is the cold calibration temperature (˚C) measured in Step 3.
8.
Send the following commands to save calibration and lock out calibration:
:CAL:PROT:CARD1:SAVE
:CAL:PROT:CARD1:LOCK
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...