1-16
Performance Verification
4.
Source each of the voltages summarized in
Table 1-7
, and verify that the temperature
readings are within limits. Be sure to select the appropriate thermocouple type for each
group of readings. (See step 2 above.)
RTD temperature
1.
Connect the precision decade resistance box (listed in
Table 1-1
) to the Model 2700
INPUT and SENSE jacks using four-wire connections. (See
Figure 1-5
for similar con-
necting scheme.) Be sure the INPUTS switch is in the FRONT position.
2.
Configure the Model 2700 temperature function for ˚C units and RTD temperature sen-
sor (
α
=0.00385) as follows:
a.
Press SHIFT then SENSOR, and note the unit displays the temperature units:
UNITS: C.
b.
Press ENTER, and note the unit displays the sensor type: SENS: TCOUPLE.
c.
Using the cursor and range keys, set the display as follows: SENS: 4W-RTD.
d.
Press ENTER, and note the unit displays: TYPE: PT100.
e.
Using the cursor and range keys, set the unit for the following display: TYPE:
PT385.
f.
Press ENTER to complete the temperature configuration process.
3.
Select the temperature function by pressing the TEMP key.
Table 1-7
Thermocouple temperature verification reading limits
Thermocouple type
Applied DC voltage* Reading limits (1 year, 18°C to 28°C)
J
K
-7.659mV
0mV
42.280mV
-5.730mV
0mV
54.138mV
-190.2 to -189.9°C
-0.2 to +0.2°C
749.8 to 750.2°C
-190.2 to -189.8°C
-0.2 to +0.2°C
1349.8 to 1350.2°C
* Voltages shown are based on ITS-90 standard using 0°C reference junction temperature. See text for pro-
cedure to set reference junction temperature.
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...