Performance Verification
1-13
Verifying resistance
Check resistance by connecting accurate resistance values to the Model 2700 and verifying
that its resistance readings are within the specified limits.
CAUTION
Do not apply more than 1000V peak between front terminals INPUT HI
and LO or more than 350V peak between SENSE HI and LO, or instru-
ment damage could occur.
Follow these steps to verify resistance accuracy:
1.
Using shielded, Teflon-insulated or equivalent cables in a 4-wire configuration, connect
the Model 2700 INPUT and SENSE jacks to the calibrator as shown in
Figure 1-5
. Be
sure the INPUTS switch is in the FRONT position.
Figure 1-5
Connections for Model 2700 resistance verification (100
Ω
to 10M
Ω
ranges)
2.
Set the calibrator for 4-wire resistance with external sense on.
3.
Select the Model 2700 4-wire resistance function by pressing the
Ω
4 key, then choose
the SLOW integration rate with the RATE key.
4.
Set the Model 2700 for the 100
Ω
range, and make sure the FILTER is on. Enable
OCOMP (offset-compensated ohms) for 100
Ω
range verification. (Press SHIFT then
OCOMP.)
5.
Recalculate reading limits based on actual calibrator resistance values.
Output
LO
Input
LO
Input
HI
Resistance Calibrator
Sense HI
Sense LO
Output
HI
Sense HI
Sense LO
Note: Use shielded low-thermal cables to
minimize noise. Enable or disable
calibrator external sense as indicated
in procedure.
Model 2700
Model 2700
Multimeter / Data Acquisition System
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
GPIB
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...