Verifying AC current
Check AC current accuracy by applying accurate AC voltage current at specific frequencies
from the AC current calibrator to the Model 2700 input, and verifying that the displayed read-
ings fall within specified limits. Follow these steps to verify AC current:
1.
Connect the Model 2700 AMPS and INPUT LO jacks to the calibrator as shown in
Figure 1-4
. Be sure the INPUTS switch is in the FRONT position.
Figure 1-4
Connections for Model 2700 AC current verification
2.
Select the AC current function by pressing the
ACI
key.
3.
Set the Model 2700 for the 1A range.
4.
Source 1A and 3A, 1kHz full-scale AC currents as summarized in
Table 1-5
, and verify
that the readings are within stated limits.
Table 1-5
ACI limits
ACV range Applied AC voltage
Reading limits @ 1kHz (1 year, 18°C to 28°C)
1A
3A
1.000000A
3.00000A*
0.99860 to 1.00140A
2.9937 to 3.0063A
* If the Fluke 5725A amplifier is not available, apply 2.2A from the calibrator. Reading limits for 2.2A are
2.1949 to 2.2051A.
Output HI
Output
LO
Input
LO
Calibrator (Output AC Current)
Amps
Model 2700
Model 2700
Multimeter / Data Acquisition System
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
GPIB
Содержание 2700
Страница 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Страница 16: ......
Страница 17: ...1 PerformanceVeri cation...
Страница 48: ...1 32 Performance Verification...
Страница 49: ...2 Calibration...
Страница 72: ...2 24 Calibration...
Страница 73: ...3 RoutineMaintenance...
Страница 79: ...Routine Maintenance 3 7 Figure 3 4 Model 7702 amps fuses CH42 Fuse CH41 Fuse Model 7702 TE122 TE121...
Страница 81: ...4 Troubleshooting...
Страница 99: ...5 Disassembly...
Страница 106: ...5 8 Disassembly...
Страница 107: ...6 ReplaceableParts...
Страница 121: ...A Speci cations...
Страница 136: ...A 16 Specifications...
Страница 137: ...B CalibrationReference...
Страница 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Страница 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Страница 173: ......
Страница 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...