62
Cables which have been shown to comply with IEC 60601-1-2:
Cable type
Shielded
Length [m]
Ferrite
Used for
PA
No
> 3
No
Potential equalization
SCB Cable
Yes
>
1
No
connecting SCB network, other SCB units
Power Cord
No
3
No
Power supply
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The ENDOFLATOR
®
40 model UI400 is intended for use in the electromagnetic environment speci
ǫ
ed below.
The customer or user of the ENDOFLATOR
®
40 model UI400 should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The ENDOFLATOR
®
40 model UI400 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The ENDOFLATOR
®
40 model UI400 is suitable for use in all establishments including
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
Ǭ
uctuations/
Ǭ
icker
emissions
IEC 61000-3-3
Complies
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Electromagnetic
Compatibility (EMC)
information
ƃƚƐƗƐƝƘƯ
Ɵƞ
ƭƛƕƚƢƠƞƜƐƓƝƘƢƝƞƙ
ơƞƒƜƕơƢƘƜƞơƢƘ
(
ƍżƁ
)