Note:
When the electrical safety test device is connected to the AUTOCON
®
II 400, an error message
will appear. This can be ignored.
The electrical safety is tested in accordance with IEC 60601-1:
– Protection class: I
– Classification: CF
– Permanently connected device: no
– Applied part with 4 function assemblies
Limit values:
– Protective ground resistance:
0.20
(with power cord)
– Earth leakage current:
–
N.C.
0.50 mA
–
S.F.C.
1.00 mA
– Housing leakage current:
–
N.C.
0.10 mA
–
S.F.C.
0.50 mA
– Patient leakage current:
–
N.C.
0.01 mA
–
S.F.C.
0.50 mA
– Patient leakage current:
–
S.F.C.
0.05 mA
(line voltage on applied part)
– Record all the data in the test report, see Chapter 7. Appendix.
– Remove all test cables.
4.4.1 Testing protective conductors
4.4.1.1 Testing protective conductor contact power
socket against housing
Test the following items:
– Housing base, see Figure 4.10
– Cover, see Figure 4.11
– Front panel, see Figure 4.12
– Front frame, see Figure 4.13
– PE contact feedthrough (3-pole), see Figure 4.14
Figure 4.10
12.06 / V1.0
4-13
AUTOCON
®
II 400
MODEL 205352 20-1
Содержание 205352 20-1
Страница 8: ......
Страница 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x ...
Страница 10: ......
Страница 13: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Страница 14: ...IV W R Q Geräteabbildungen Photographs of the equipment Imágenes del equipo I O Y U T E ...
Страница 146: ......
Страница 152: ......
Страница 156: ......
Страница 158: ......
Страница 172: ......
Страница 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1 ...
Страница 192: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......