
12.06 / V1.0
3-11
AUTOCON
®
II 400
MODEL 205352 20-1
Error code
Description of error
Solution
Priority
I-81
Short circuit in instrument interface
Arrange for Black Box to be tested and
replaced if necessary
Low
I-82
Short circuit in activation recognition
Arrange for Black Box to be tested and
replaced if necessary
Low
I-83
Cut-off in activation recognition
Arrange for Black Box to be tested and
replaced if necessary
Low
I-84
Instrument interface: R-Window
Arrange for Black Box to be tested and
replaced if necessary
Low
I-85
Short circuit in activation recognition
Arrange for Black Box to be tested and
replaced if necessary
Low
I-86
Illegal activation type
Arrange for Black Box to be tested and
replaced if necessary
Low
I-87
Framing error
Arrange for Black Box to be tested and
replaced if necessary
Low
I-88
Parity error
Arrange for Black Box to be tested and
replaced if necessary
Low
I-89
Unknown ECB identifier
Arrange for Black Box to be tested and
replaced if necessary
Low
I-8A
CRC error in instrument recognition
Arrange for Black Box to be tested and
replaced if necessary
Low
I-8B
Instrument interface: R_S3
Check connected instrument and
replace if necessary
Low
I-8C
Non permissible socket module
recognized
Arrange for Black Box to be tested and
replaced if necessary
Low
I-8D
Instrument interface: R_S4
Check connected instrument and
replace if necessary
Low
I-8E
Instrument interface: R_S8
Check connected instrument and
replace if necessary
Low
I-90
EEPROM write error
Arrange for Black Box to be tested and
replaced if necessary
Low
N-01
Interrupt error
Arrange for Black Box to be tested and
replaced if necessary
Low
N-02
CRC error following a reset
Arrange for Black Box to be tested and
replaced if necessary
Low
N-03
+5 V too low
Arrange for Black Box to be tested and
replaced if necessary
Low
N-04
+5 V too high
Arrange for Black Box to be tested and
replaced if necessary
Low
N-05
Undervoltage of ±12 V
Arrange for Black Box to be tested and
replaced if necessary
Low
N-06
Overvoltage of ±12 V
Arrange for Black Box to be tested and
replaced if necessary
Low
N-07
Current for NE monitor absent
Arrange for Black Box to be tested and
replaced if necessary
Low
N-08
SPI receive error
Arrange for Black Box to be tested and
replaced if necessary
Low
N-09
Stack error
Arrange for Black Box to be tested and
replaced if necessary
Low
N-0D
Error in state variable
Arrange for Black Box to be tested and
replaced if necessary
Low
N-0E
CRC error during operation
Arrange for Black Box to be tested and
replaced if necessary
Low
N-10
Defective current sensor 1
Arrange for Black Box to be tested and
replaced if necessary
Low
N-11
Defective current sensor 2
Arrange for Black Box to be tested and
replaced if necessary
Low
Caution:
Always unplug the device from the power supply before carrying out any
repairs on it!
Содержание 205352 20-1
Страница 8: ......
Страница 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x ...
Страница 10: ......
Страница 13: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Страница 14: ...IV W R Q Geräteabbildungen Photographs of the equipment Imágenes del equipo I O Y U T E ...
Страница 146: ......
Страница 152: ......
Страница 156: ......
Страница 158: ......
Страница 172: ......
Страница 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1 ...
Страница 192: ......
Страница 198: ......
Страница 200: ......
Страница 202: ......
Страница 203: ......