TECHNICAL DESCRIPTION
ULTRAFLOW
®
54 DN150-250
5512-876 GB/04-2012/Rev. F1
54
10.4.2
Flow meter adjustment
Open ”Flow Meter Adjustment”:
”Read from Meter”: Reads data from the connected flow sensor.
Flow curve number - 5945357 -
and meter dimensions appear from the heading. This number
will also appear from the meter’s label.
The field “Flow Curve” shows the values of the sensor in question compared to the standard
curve. These values are also shown in the form of a graph.
The required correction of q
i
, 0.1xq
p
and q
p
can be entered into the field ”Flow Curve
Correction”.
”Write to Meter”:
Writes the correction to the connected flow sensor.
After the adjustment the flow sensor is ready for renewed test.
10.4.3
Programming of standard flow curve
Open ”Flow Meter Adjustment”:
The 59xxxxx no. appears from
the sensor's type label.
”Read from database”:
Enters the selected standard flow curve into the program.
”Write to Meter”:
Programs the flow sensor with the entered standard flow curve.
The flow sensor is now ready for test.