TECHNICAL DESCRIPTION
ULTRAFLOW
®
54 DN150-250
5512-876 GB/04-2012/Rev. F1
53
10.3.4
Help
About:
Includes program numbers and revisions of the
various components of the installed version.
10.4
Application
Flow sensor adjustment.
Before adjusting a sensor you must make sure that the sensor operates satisfactorily in the flow rig in question.
See
paragraph 9 Calibrating ULTRAFLOW
®
If it is necessary to adjust the sensor more than a few percent, the sensor is probably defective, or has a different
error, and should not be adjusted.
10.4.1
COM-port selection
Open ”Select Com Port”:
Select a COM-port for ULTRAFLOW
®
X4.
The USB driver must be installed before connecting the interface.
The related COM port will not appear from the list until the USB
interface has been connected.
Select COM port for Pulse Divider.
Activate ”OK” in order to save the selected ports.