TECHNICAL DESCRIPTION
ULTRAFLOW
®
54 DN150-250
5512-876 GB/04-2012/Rev. F1
48
9.4.5
Spare parts
Description
Type No.
Battery D-cell
66-00-200-100
Cable retainer (secures the battery)
1650-099
2-pole plug (female)
1643-185
3-pole plug (female)
1643-187
PCB (66-R)
5550-517
Table 18
9.4.6
Battery replacement
If the PULSE TESTER is used continuously we recommend that the battery be replaced once a year.
Connect the battery to the terminals marked ”Batt”, the red wire to + and the black one to -.
Current consumption:
Power consumption with no sensors connected
400
µ
A
Max. power consumption with two ULTRAFLOW
®
connected
1.5 mA
Note: If the base unit is fitted with battery or externally supplied, the PULSE TESTER’s integral supply must be
disconnected (the plug must be removed).