TECHNICAL DESCRIPTION
ULTRAFLOW
®
54 DN150-250
5512-876 GB/04-2012/Rev. F1
41
9
Calibrating ULTRAFLOW
®
Calibration can be based on:
•
Pulses in standard mode
•
Pulses in test mode
•
Pulses using PULSE TESTER type 66-99-279
•
Serial data with the sensor in test mode
9.1
Test instructions for ULTRAFLOW
54, DN150-250
9.1.1
Installation angle for ULTRAFLOW
®
ULTRAFLOW
®
54 DN150-250
9.1.2
Technical data
Supply:
3.6 VDC ±0.1 V
Pulse duration:
2…100 ms (depending on programming)
Pulse figure:
Depends on programming. Appears from type label.
Start-up:
It takes 16 seconds from start-up until true flow reading has been obtained and calibration
can start.
ULTRAFLOW
®
can be installed
horizontally, vertically, or at an angle.
Figure 23