TECHNICAL DESCRIPTION
ULTRAFLOW
®
54 DN150-250
5512-876 GB/04-2012/Rev. F1
38
Flow sensor tolerances qi:qp 1:100 (qp 150 m³/h)
0
1
2
3
4
5
6
1
10
100
1000
Flow [m³/h]
T
o
le
ra
n
c
e
s
[
%
]
EN1434 cl.3
EN1434 cl.2
½ EN1434 cl.2
qs
qp
0,1x qp
qi
Diagram 14
8.11
Power consumption
The power consumption of ULTRAFLOW
®
is as follows:
Max. average
50 µA
Max. current
7 mA (max. 40 ms)
8.12
Interface plug/serial data
ULTRAFLOW
®
54 is fitted with a four-pole connector under the cover. The cover is supplied with a factory seal, and
in connection with verified sensors it will be a laboratory seal (legal seal). Thus, it is not possible to access the
connector without breaking the seal.
The connector is used for:
•
Meter programming, including adjustment of correction curve by means of METERTOOL
•
Bringing the sensor into test mode
•
Reading accumulated water quantity in connection with calibration
•
External control of start/stop in connection with calibration
The interface connector is constructed as shown in
Figure 20.