7 Mounting
22
7.2.2
Installation in a pipeline with G 2" union nut
1. Install the fitting 406090 into the pipeline,
see chapter 7.2.1 "Recommendation for installing into
2. Check that the seal (7) is on the device (8).
3. Place the union nut (4) on the fitting (6).
4. Place the circlip (3) into the groove (5).
5. Position the device so that the arrow on the side of the housing is pointing in the direction of flow (1):
the counters are incremented.
6. Place the device (8) into the fitting (6).
7. Manually tighten the union nut (4) on the device (8).
NOTE!
Please refer to the recommended installation type!
See chapter 7.2 "Installation into the pipeline", page 20 and operating manual "Fittings for flow rate
sensors" (406090)
NOTE!
To ensure more accurate measurements and good stability of the flow rate zero point, bring the sensor
into contact with the measurement medium at least 24 hours prior to calibration.
(1)
Flow direction
(2)
Measuring electrodes
(3)
Circlip
(4)
Union nut
(5)
Groove
(6)
Fitting 406090
(7)
Seal
(8)
Device