8 Startup and parameterization
42
Determining the fitting's K-factor with a Teach-In process in relation to a volume (TEACH V.)
1. If you do not wish to change any further parameters, go to the END function in the parameters menu
and press the
key in order to save the settings and return to the process level.
NOTE!
The device starts using the new K-factor as soon as the function SAVE YES is confirmed when leaving
the parameters menu.
1. Prepare a container with a known volume.
2. Stop medium circulation.
3. Confirm TEACH V.: FILL. END is displayed.
K-FACTOR
K = 2.8500
The display shows the K-factor for the fitting (406090)
used, whether entered or determined using Teach-In.
TEACH V.
FILL. END
1. Charge the pipe to fill the tank.
2. When the tank is full, confirm FILL. END.
0000.0 L
TEACH F.
1. Enter the volume (value between 0.1 and
9999.9) of medium that flowed past the device
and confirm.
K = 2.9000
The device calculates the K-factor of
the fitting and displays it.
1. Confirm the displayed value.
OUTPUT
RETURN
VALID N/Y
1. Confirm the K-factor determined through Teach-
In or not.