9 Maintenance
78
9.5.3
Solving a problem without a warning or error message with a green device status
LED
Device sta-
tus LED
Analog out-
put AO1
Digital outputs
DO1/DO2/DO3
Meaning
Measure
Green
4 to 20 mA
Depending on
switching thresh-
olds or switched
a
The device is not
measuring the flow
rate correctly.
•
Check that the K-factor matches the
fitting used
•
Check that the KW value has not been
switched
•
Perform a Teach-In process to deter-
mine the K-factor for the fitting used
Green
4 to 20 mA
Depending on the
switching thresh-
olds or switched
The displayed flow
rate is not zero
even though the
flow rate in the pipe-
line is zero.
•
Check that the flow rate in the pipeline
is zero-
•
Check that there are no air bubbles in
the pipeline-
•
Check the filter level-
•
Calibrate the flow rate zero point.
Green
4 to 20 mA
Depending on the
switching thresh-
olds or switched
The displayed flow
rate is always zero.
•
Check that the flow rate in the pipeline
is not zero.
•
Check that the K factor or KW value
are not too low.
•
Check that the measuring electrodes
are mounted so that they are perpen-
dicular to the direction of flow.
•
Select a smaller flow rate unit or in-
creased the number of decimal places
displayed.
Green
4 to 20 mA
Depending on the
switching thresh-
olds or switched
The displayed flow
rate is not stable.
•
Check whether the medium is flowing
in the pipeline
•
Select a higher filter
Green
4 to 20 mA
Depending on the
switching thresh-
olds or switched
The flow rate dis-
play is changing at
a very slow place.
•
Check whether the medium is flowing
in the pipeline
•
Select a smaller filter
a
If the output has been configured to switch upon generation of a warning,
see chapter 8.6.5 "Configuring out-